INSTRUCTIONS FOR USE
Reditest Panel-Dip on-site drug screening devices are easy to use! Please follow the simple instructions below.
NOTE: ALL POSITIVE RESULTS ARE PRESUMPTIVE AND SHOULD BE CONFIRMED BY AN ALTERNATIVE METHOD (e.g. GC/MS)
STEP 1
Remove device from foil pouch.
STEP 2
Pull plastic cap from device.
STEP 3
Dip revealed sample pads into urine sample for a minimum of 15-30 seconds until urine is seen wicking into the reading zone. Make sure each pad is dipped into the specimen to the level of the wavy lines without touching the plastic card to the urine.
STEP 4
Remove the device from the urine and replace the plastic cap. Place on a flat surface.
STEP 5
Read test results at 5 minutes. (Do not interpret results after 4 hours as false results may occur. Note: Any line at all in the test zone is to be interpreted as negative.)
NEGATIVE
Colored line appears
in Test (T) ZonePOSITIVE
NO Colored line in
Test (T) ZoneINVALID
Colored line must
appear in Control (C)
Zone for test to
be valid
LABORATORY URINE DRUG TESTING
If test is positive, apply security seal over the lid and agency label with test(s) requested around the bottle. Send bottle and contents to laboratory for confirmation.
Sending Presumptive Positive Specimens to the laboratory for GC/MS Confirmation
- Perform the on-site test according to the instructions provided above.
- Complete an Urine Drug Test Lab Requisition Form and Label to indicate which presumptive positive drugs you want to confirm through the lab. Make sure the donor signs the form. NOTE: If you wish to document preliminary positives and negatives for your own record, complete a Preliminary Result Form.
- Secure the positive specimen in the bottle. To prevent leakage, make sure the lid is threaded properly and tightly fastened.
- Place the security seal over the top of the bottle.
- Fill out the Specimen ID/Chain of Custody Label and indicate the positive drug(s) on the line marked "GC/MS Confirm:___________".
- Place the provided label on the bottle.
- Instruct the donor to initial the security seal.
- Seal the bottle in a plastic baggie with the absorbent material.
- Include the “Lab Copy” of the Test Requisition Form in the shipping bag (five or more specimens) or the mailer box (fewer than five specimens) with the specimen.
- Send to RTL for GC/MS confirmation.
NOTE: If your account was not set up to include confirmation request forms, please contact your sales representative.
*PLEASE DO NOT SEND NEGATIVE SPECIMENS TO RTL FOR CONFIRMATION. NEGATIVE SPECIMENS WILL NOT BE CONFIRMED.